TCi Operator Manual
59
Step 3 - Create the calibration method.
The calibration method contains the timing parameters, power level, and calibration material
group to be used for the calibration.
Building the Calibration
More information on calibration building can be found in the Acquire Calibration Data and
Compute Results sections.
Step 1 - Acquire calibration data.
Test all calibration materials at the desired temperature.
Step 2 - Compute results.
Remove any outliers from the calibration test data and plot the calibration curve.
Using the Calibration
Step 1 - Create a test method that uses the calibration method created for the user
calibration.
Step 2 - Perform a test.
6.4
Create Calibration Materials
Step 1 – Open the Calibration Material table.
Step 2 – Click the Add button.
Step 3 – Enter the Calibration Material name.
Step 4 – Enter a material description (optional).
Step 5 – Click the Add Row button ( + button) twice.
Step 6 – Enter a thermal conductivity value.
Step 7 – Enter a thermal effusivity value.
Step 8 – Click the Save button.
Step 9 – Repeat steps 2-8 as needed.
6.5
Create Calibration Material Group
Step 1 – Open the Calibration Material Group table.
Step 2 – Click the Add button.
Step 3 – Enter the Calibration Material Group name.
Step 4 – Select a Diagnostic Material from the list of calibration materials.