TCi Operator Manual
104
C
m
m
T
/
1
1
'
in
V/
s
(31)
Is the corrected slope for temperature T.
For example, suppose a material under test has
1/m
25
=150
s/V
and
1/m
70
=160
s/V
, and the
k
calibration has these parameters:
Slope
=100
smK/WV
,
Intercept
=90
s/V
and
m
*
=0.0015
V/
s
.
Calculation of
k
at 25
C is:
035
.
1
100
90
55
.
193
100
90
005166
.
0
1
100
90
0015
.
0
150
/
1
1
1
*
25
Slope
Intercept
m
m
k
W/mK
.
Calculation of
k
at 70
C is:
From the
1/m
Vs. temperature graph above,
C
25
=54 and
C
70
=62. Hence
C
=+8
s/V
.
152
1
8
160
1
/
1
1
/
1
1
'
70
C
m
C
m
m
T
0685
.
1
100
90
85
.
196
100
90
00508
.
0
1
100
90
0015
.
0
152
/
1
1
1
*
'
70
Slope
Intercept
m
m
k
W/mK
.
Without correction the calculated
k
would be:
20
.
1
100
90
52
.
210
100
90
00475
.
0
1
100
90
0015
.
0
160
/
1
1
1
*
70
Slope
Intercept
m
m
k
W/mK
.
Notes:
1.
For
k
calculation, the
C(T)
must correspond to the applicable power level.
2.
This correction is most effective for low
k
materials. For high
k
materials, the correction
may be negligible.
9.7
Error analysis
9.7.1 General
There are in general 2 types of errors (sometimes called uncertainty) – offset and variation. Offset
errors affect the accuracy of the measurement while the variation affects the repeatability.