TCi Operator Manual
52
The count is the number of measurements to be taken on each calibration material.
Step 22 – Select the first calibration material.
Step 23 – Place the calibration material.
Solid non-porous materials: Place three drops of contact agent on the sensor surface. Place the
calibration material on the sensor. Place the 500 gram weight on the sample. The weight may
be omitted if the sample weighs more than 150 grams.
Solid porous samples: Apply Wakefield 120 thermal grease to the sensor. Place the calibration
material on the sensor. Place the 500 gram weight on the sample. The weight may be omitted if
the sample weighs more than 150 grams.
Liquid samples: Fill the 50 mL beaker to 35 mL. Invert the sensor and slide it into the liquid at an
angle to prevent bubbles from forming on the sensor surface. Alternately, the small volume test
kit may be used.
Step 24 – Click the Start button.
Step 25 – Repeat steps 22 to 24 for each other calibration material.
Test a minimum of three calibration materials.
Note: Examine results for bad measurements. If one measurement is very high, de-select
it. It is also recommended to take another set of measurements on that calibration
material.
Step 26 – Click the Next button.