8 System Calibration and Maintenance
8.1
Calibration
C-Therm personnel conduct factory calibrations and the system is shipped ready for use.
Multiple calibrations can be included with different test times and power levels, depending on the
requirements of the user.
The calibration can be confirmed by performing a reference material test with Distilled Water, LAF
6720, Pyrex, Pyroceram, or Phosphor Bronze. Refer to the Reference Material Test section for
instructions on performing a reference material test.
User calibration capabilities are available with the Max K Module. See the User Calibration
chapter for details.
8.1.1 Unlock Calibrations/User Calibration
Note: Factory calibrations are loaded on the sensor and do not require import. They do,
however, need to be unlocked in the software. An authorization XML file is provided on
the C-Therm TCi Software CD and will unlock the modules purchased. The file can also be
obtained by emailing [email protected].
Step 1 – Select Authorization from the Tools menu.
Step 2 – Click the Open File button.
Step 3 – Locate the Authorization TH89-05-XXXXX.xml file.
The file is located on the C-Therm TCi Software CD.
Step 4 – Select the XML file.
Step 5 – Click the Open button.