TCi Operator Manual
86
8.2.2 System Checks
Step 1 – Open the diagnostics window.
Click the Diagnostics button on the toolbar.
The first tab, labeled Instrument Status, contains information about the instrument.
Step 2 – Click the Read Sensor Temperature button.
The reading is displayed in the ambient temperature column of the table.
Step 3 – Click the Calibration Data tab.
Step 4 – Select a calibration method.
The calibration method drop-down list is located at the top of the diagnostics window.
Calibration results for the selected calibration method are displayed in the top table. All
calibration results on the sensor ID chip are listed in the bottom table. If a calibration is labeled
as LOCKED, it must be opened with an authorization key from C-Therm Technologies.