TCi Operator Manual
83
*Note: Due to variability between batches, there are multiple calibration material records
for LAF 6720 and Phosphor Bronze. When performing a test, select the record matching
the label on the reference sample. All reference material values are located in Appendix D.
Step 1 – Prepare the reference material and sensor.
Distilled Water - Pour 35 mL of water into a 50 mL beaker. Slowly slide the sensor surface into
the water at an angle to prevent bubbles from forming on the sensor surface.
LAF 6720 - Place the sample on the sensor and place a weight on the sample to ensure good
contact.
Pyrex, Pyroceram, and Phosphor Bronze - Place the required amount of contact agent on the
sensor surface (-20 to -5 °C- three drops of Glycol, 5 to 70 °C three drops of water, 70 °C and
higher, or for long term tests Wakefield 120 thermal grease). Place the sample on the sensor and
place the weight on the sample to ensure good contact.
Step 2 - Click Reference Material Test from the Tools menu.
Leave the Perform New Test option selected.
Step 3 - Click the Next button.