Operation
E164809_2_002
37 / 72
A flag in the
Info
Toolbar indicates a calibrated minispec application.
An uncalibrated minispec application is labeled with the following flag:
The Detect raw Signal Command
In some cases, for example to adjust the acquisition parameters, it might be helpful to detect
the signal without data processing. This command was implemented to execute only the
pulse sequence and acquisition without data processing or result output.
Shortcut: Keys: ALT+S
The Run Repetitive Command
With this command a measurement can be repeated several times at a regular interval.
Shortcut Keys: ALT+R
Select
'Repetitive Measurement Settings
' from the
Parameter
menu to determine the
waiting time in seconds before a repetitive minispec measurement, the delay between two
successive repetitions, and the maximum number of repetitive measurements.
6.2.4.6
The Parameter Menu
All parameters for acquisitions, applications and repetitive measurements can be defined with
the assistance of the commands in this menu:
Figure 6.9: The Parameter Pull-Down Menu
Acquisition parameters are application dependent values (for example receiver gain, recycle
delay and number of scans to accumulate) that help define data acquisition. These values
can be modified when a minispec is connected to the PC and a minispec application has
been loaded. To modify acquisition parameters open the
Acquisition Parameter Table
.
Shortcuts: Standard Tool Bar:
Keys: ALT+P
To save the modifications to file, click <Save> before you leave the
Parameter Table
.
Note:
Calibrated minispec applications will loose their calibration, if parameters are changed.
If the
Save
button stays disabled on a connected minispec, customize your minispec
software to provide access to the calibrated application.
Summary of Contents for minispec LF Series
Page 1: ...minispec minispec LF series User Manual Version 002 Innovation with Integrity AIC...
Page 8: ...Introduction 8 72 E164809_2_002...
Page 13: ...Safety E164809_2_002 13 72 Figure 2 2 Example of a Stray Field Plot for the LF90 Top View...
Page 14: ...Safety 14 72 E164809_2_002...
Page 18: ...Transport Packaging and Storage 18 72 E164809_2_002...
Page 50: ...Operation 50 72 E164809_2_002...
Page 56: ...Maintenance 56 72 E164809_2_002...
Page 58: ...Replacement of Parts 58 72 E164809_2_002...
Page 64: ...Technical Data 64 72 E164809_2_002 Figure 12 1 An Example of a minispec Rating Plate...
Page 66: ...Contact 66 72 E164809_2_002...
Page 68: ...List of Figures 68 72 E164809_2_002...
Page 71: ...E164809_2_002 71 72...
Page 72: ...Bruker Corporation info bruker com www bruker com Order No E164809...