Operation
E164809_2_002
33 / 72
The Instrument Settings Command
Minispec instrument settings are instrument dependent values such as the pulse lengths and
the magnetic field at NMR resonance. Instrument settings differ from minispec to minispec,
and must be tuned the first time a minispec is connected. You can use this command to
change the settings of the instrument with manual input.
Figure 6.7: The minispec Settings Table
You can change the listed settings by manual input or with the Update Settings command.
Grayed entries are disabled.
Shortcut:
Standard
Tool Bar:
The settings that can be changed are shown in the figure above. If the
minispec Settings
are disabled on a connected minispec, customize your minispec software (through the
Welcome Box) to access the calibration parameters.
The
Advanced
button in the settings table allows access to a list of instrument setting table
names. The user may select the appropriate settings table to be used by the instrument,
define custom names for the settings tables with the help of the
Customize
button (e.g.
settings table name could be set to the probe name), and save the current entries under a
new name.
Entries in the settings table about SFC are only relevant in the case of the SFC-Analyzer. If
the SFC-item in the upper middle part of the settings table is marked, then the
Update
Settings
and the
Daily Check
routines will include steps to determine the SFC-specific
values. The factors (K, F, and O) and the sampling windows for a solid fat content analysis
are definable via the
SFC-Settings
button.
Summary of Contents for minispec LF Series
Page 1: ...minispec minispec LF series User Manual Version 002 Innovation with Integrity AIC...
Page 8: ...Introduction 8 72 E164809_2_002...
Page 13: ...Safety E164809_2_002 13 72 Figure 2 2 Example of a Stray Field Plot for the LF90 Top View...
Page 14: ...Safety 14 72 E164809_2_002...
Page 18: ...Transport Packaging and Storage 18 72 E164809_2_002...
Page 50: ...Operation 50 72 E164809_2_002...
Page 56: ...Maintenance 56 72 E164809_2_002...
Page 58: ...Replacement of Parts 58 72 E164809_2_002...
Page 64: ...Technical Data 64 72 E164809_2_002 Figure 12 1 An Example of a minispec Rating Plate...
Page 66: ...Contact 66 72 E164809_2_002...
Page 68: ...List of Figures 68 72 E164809_2_002...
Page 71: ...E164809_2_002 71 72...
Page 72: ...Bruker Corporation info bruker com www bruker com Order No E164809...