Operation
48 / 72
E164809_2_002
6.3.1
Daily Check
As a first action, run the
Daily Check
function to validate the system. The software will then
guide you through the procedure, requesting the Daily Check sample. Insert the
corresponding samples with care into your NMR system.
Finally, the test should conclude, and the instrument should be valid for 24 hours. In case
there is a Daily Check failure, it is recommended to run the
Update Settings
procedure and to
follow the corresponding instructions. In any case if the
Update Settings
routine does not run
automatically, the Daily Check procedure has to be repeated.
During Daily Check, the system runs a number of tests, all following one after the other.
If all the tests passed, the software will indicate this accordingly. The corresponding message
will also be displayed in the main menu.
Summary of Contents for minispec LF Series
Page 1: ...minispec minispec LF series User Manual Version 002 Innovation with Integrity AIC...
Page 8: ...Introduction 8 72 E164809_2_002...
Page 13: ...Safety E164809_2_002 13 72 Figure 2 2 Example of a Stray Field Plot for the LF90 Top View...
Page 14: ...Safety 14 72 E164809_2_002...
Page 18: ...Transport Packaging and Storage 18 72 E164809_2_002...
Page 50: ...Operation 50 72 E164809_2_002...
Page 56: ...Maintenance 56 72 E164809_2_002...
Page 58: ...Replacement of Parts 58 72 E164809_2_002...
Page 64: ...Technical Data 64 72 E164809_2_002 Figure 12 1 An Example of a minispec Rating Plate...
Page 66: ...Contact 66 72 E164809_2_002...
Page 68: ...List of Figures 68 72 E164809_2_002...
Page 71: ...E164809_2_002 71 72...
Page 72: ...Bruker Corporation info bruker com www bruker com Order No E164809...