Operation
36 / 72
E164809_2_002
Measurement Tool Bar:
The command
Continue
is used to leave an intermediate process in order to continue
running the minispec application (e.g. during a waiting time, a data acquisition or a fit
display).
Shortcuts:
Standard Tool Bar:
Measurement Tool Bar:
Keys: ESC
The Calibrate Command
A typical minispec application detects NMR signals of the sample and then calculates the
result from the NMR signal. In some cases, the result can be directly calculated from the
measured signal (e.g. a relaxation time). In other cases the minispec applications may use a
reference method and calibration line to determine a sample property (e.g. oil, water or fat
content) from the NMR signal. As a consequence, the NMR signal must first be calibrated
with respect to sample properties. Select
Calibrate Sample
from the
Measurement
menu.
Shortcuts: Standard Tool Bar:
Measurement Tool Bar:
Keys: ALT+C
Running a minispec sample calibration leads the user through several steps. In most cases
some application specific values (see: The Application Parameter Table Command) need to
be defined.
To calibrate a minispec application, proceed as follows:
1. Prepare at least 3 samples of known content for measurement.
2. Perform test measurements with each calibration sample to adjust the acquisition
parameters from the
Acquisition Parameter Table
in the
Parameter
menu. Adjust the
receiver gain to the sample with the strongest signal.
3. Save the parameters and leave the parameter table.
4. Run the sample calibration and follow the prompts.
Summary of Contents for minispec LF Series
Page 1: ...minispec minispec LF series User Manual Version 002 Innovation with Integrity AIC...
Page 8: ...Introduction 8 72 E164809_2_002...
Page 13: ...Safety E164809_2_002 13 72 Figure 2 2 Example of a Stray Field Plot for the LF90 Top View...
Page 14: ...Safety 14 72 E164809_2_002...
Page 18: ...Transport Packaging and Storage 18 72 E164809_2_002...
Page 50: ...Operation 50 72 E164809_2_002...
Page 56: ...Maintenance 56 72 E164809_2_002...
Page 58: ...Replacement of Parts 58 72 E164809_2_002...
Page 64: ...Technical Data 64 72 E164809_2_002 Figure 12 1 An Example of a minispec Rating Plate...
Page 66: ...Contact 66 72 E164809_2_002...
Page 68: ...List of Figures 68 72 E164809_2_002...
Page 71: ...E164809_2_002 71 72...
Page 72: ...Bruker Corporation info bruker com www bruker com Order No E164809...