NMR Parameters
54 / 72
E164809_2_002
8.4
Acquisition Pre-Delay Time
In some cases a defined delay is desired between the time a sample is inserted and the time
of measurement start (e.g. if time is needed to allow the temperature of the sample to
equilibrate in the magnet). The
Acquisition Pre-Delay Time
option allows such a delay to be
set (in seconds). The waiting time is added to the recycle delay of the first scan of the data
acquisition.
8.5
Dummy Shots
Pulses before the measurement can be made prior to the experiments (0-32000), in order to
ensure that the first scan has the same history as all subsequent scans.