Planning details
6.5.6
Connection and assembly of electrical distributor
The assembly and the connection of the electrical power distributor to the main power supply is
the MR system owner's task. In order to be able to finish the electrical installation before instal-
lation of the MR system begins, the electrical power distributor can be delivered in advance.
The electrical power distributor is assembled in a suitable position on the wall in the technical
room.
The maximum distance to the installation site of the electronics cabinets of the MR system is
limited due to the 8 m long connection lines between the electrical distributor and the electronics
cabinets.
Schematic diagram of the electrical distributor
Figure 6.17: Schematic diagram of the electrical distributor
6.6
Supervision signals
The operation state of the magnet can be remotely supervised using the remote alarm connec-
tion of the magnet monitoring unit (BMU). The connection to a remote supervising and alarm
system of the building is done via a male 4 pin Amphenol connector of type C091A 4M. The
BMU provides two different alarm signal types with the following potential free pin assignment:
1
Horizontal assembly position: approx. 1200 mm above the floor
2
Mains connection line on the building side
3
Outgoing circuits for the MR system's consumers
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...