78
ID: 182463115
Delivery and transport
Transport limitations [schreibgeschützt]
Transport limitations
Table 7.2: Transport limitations
7.5
Moving into the building
Before the MR instrument can be delivered and moved to the installation site, the load-bearing
capacity along the transport route must be calculated and verified by a qualified structural engi-
neer.
The building dimensions along the entire transport routes must be carefully checked and must
allow the transportation according to the transport requirements given below.
For that purpose, the transport weights of magnet and cabinets are provided in section
, the transport dimensions of the magnet is given in the figure below.
transport restrictions 117/16 USR
Limit value
Maximum horizontal acceleration
2.0 g
Maximum vertical acceleration
3.0 g
Maximum duration during maximum acceleration
20 ms
Maximum tilt angle
10°
No sliding on loading surfaces.
--
No contact with other transport goods
--
Only gentile placement
--
No tilting by more than 10° from vertical
--
Maximum inclination along the transport route
15 mm/m
NOTICE
Hazard of crash or worse injury to hands or feet
Make sure that the occupational standards for minimum safety distances between building
structures and heavy loads are fulfilled along the entire transport route.
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...