Planning details
If the system is installed within an HF shielded room, suitable connecters, extensions and HF
filtered feed through elements can be ordered under the
Bruker
part number T11976
.
6.10
Cold water supply
Part of the heat dissipated from the MR system is released via a heat exchanger to a cold water
supply system.
Heat dissipated to the cold water system
Heat dissipated to the cold water system
Table 6.20: Heat dissipated to the cold water system
1
Extraction tube for anesthetic gas (inner diameter of the tube=25 mm) and connection
to the white plastic duct 90° bend with inner diameter of 110 x 54 mm.
2
Set for anesthetic gas extraction system
Standard configuration
MRI CryoProbe™
option
Heat dissipated into the water
~ 11 kW
Additional ~ 8.5 kW
Supply temperature
8-12 °C (*)
8 – 24 °C
Maximum pressure (forward flow)
15 bar
7 bar
Minimum pressure difference (forward
flow)
5 bar (*)
3.2 bar
Minimum flow
45 l/min (*)
7 l/min
Recommended particle filter
500 µm
500 µm
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...