Planning details
TN-S connection diagram of the MR system (50 Hz)
Figure 6.16: TN-S connection diagram of the MR system (50 Hz)
Connection values for 117/16 USR
Connection values in the technical room
Connection values in the technical room
Table 6.12: Connection values in the technical room
The electrical power distributor provides the following connection points, which should be used
solely for operating the MR system:
1a
Equipotential bonding: Earthing point on the building side, which is used solely for the
MR system
1b
Equipotential bonding: Equipotential bonding bars in the technical room as a central
neutral point of all MR system parts
2
central Bruker BioSpin MRI electrical distributor outgoing circuits to all MR system con-
sumers
3
Generator or transformer on the building side for provided the mains voltage required
These units must be provided by the operator
Technical room
Specification
Connection lines to the electrical distributor
5 x 35 mm
2
, shielded, max. 100 m long
Nominal voltage
400/230 V, +/- 10%, 3 N~
Frequency
50 or 60 Hz
Fuse
>80 A
Max. input power in operation
22.5 kVA
Input power in stand-by operation
11.5 kVA
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...