Planning details
6.8
Lighting system
The lighting facility is not an integral part of the supply and must be finished by the customer
before installation begins.
There are no system specific requirements for the operating and technical rooms. When select-
ing the lighting facility in these areas, the country specific specifications and requirements for
work place lighting must be met (occupational standards).
In the area of the magnet room, lighting facilities must not be below the minimum distances to
an MR system listed in the chapter
Measurements and room dimensions [
. When se-
lecting lighting methods, their compliance with magnetic stray fields must be noted. The use of
combinational circuit parts, transformers, and dimmers is not permitted in the immediate vicinity
of the magnet. The operating areas in front of and behind the magnet should be very well lit.
If an MR system is installed in an HF shielded room, the magnet room lighting and any AC sock-
ets for additional equipment must be connected to the central electrical power distributor at the
connection points provided for this purpose. All supply lines into the HF shielded room must
pass through appropriate filter plates.
6.9
Ventilation and air conditioning
6.9.1
Overview
The following areas and units must be fitted with a cooling system available for operating the
MR system on the building side:
Overview of the types of cooling system
Table 6.16: Overview of the types of cooling system
(*) the real requirement depends on the configuration of the instrument
No additional ground loop should be established by the lighting system. When there is discon-
nected ground at the electrical power distributor, there must be infinitely large resistance be-
tween each ground contact of the lighting facility and all conducting points within the HF
shielded room.
Unit
Type of cooling
system
Cooling perfor-
mance
Electronics cabinets in the technical room (*)
Air conditioning unit
~ 4.5 - 15 kW
Cooling circuits for magnet and gradient
Cold water supply
~ 11,5 kW
MR system in the magnet room
Air-conditioning sys-
tem
~ 0.5 kW
Anesthetic gas extraction in the magnet room
Gas extraction
no
Operating room (*)
General room air
ventilation system
~ 0.5 kW
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...