Water supply installation in the technical room
Figure 6.22: Water supply installation in the technical room
Each connection point must be provided with a ¾"
shut-off valve
and
tube nipple
at a height
of about 1 m. The cold water system lines of the building side must have a cross-section of at
least 1" and be thermally insulated.
Bruker BioSpin MRI recommends on-site installation of a temperature and pressure probe/reg-
ulation for monitoring the properties of the cold water system going to the MR system.
If no in-house cold water system is available at the installation site, a local fixed, decentralized
cold water production system (inside or outdoors) can be included in the plans.
1
Backflow (warm)
2
Forward flow (cold)
3
Insulation to prevent condensation
4
Lock valves with 3/4" tube nipples
5
Connection to the basic system
6
MRI CryoProbe™ option: An additional parallel connection is required
7
shut-off valve
8
tube nipple
NOTICE
Water in the area of the electrical systems due to condensation
To avoid water collecting in the technical room due to condensation, the cold water pipes must
have suitable thermal insulation. Dripping water at the connections points must be captured
using appropriate collection containers.
These systems are planned, installed, and maintained by local refrigeration technology firms.
Bruker BioSpin MRI can help with planning and supplying these systems if necessary, but not
with installing and maintaining them.
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...