Planning details
Specific solutions are possible in special cases for achieving load distribution in existing build-
ings. These solutions should be analyzed in relation to their vibration behavior. Discuss the sit-
uation and possible counter-measures with the planners from your local Bruker BioSpin MRI
office.
6.4.2
Magnet foundation
The ferromagnetic density should not exceed 100kg/m
2
in the area of the magnet foundation. If
this limit cannot be maintained for new buildings, non-magnetic reinforcement materials (e.g.
high-yield austenitic steels) should be considered.
The magnet foundation must be constructed in the form of a bonded screed. If an HF shielded
room is used, the bonded screed must extend at least over the surface area of the box.
Construction of the magnet foundation, if the original floor construction at the magnet installation site is based on a floating floor.
Figure 6.11: Construction of the magnet foundation, if the original floor construction at the magnet installa-
tion site is based on a floating floor.
1
accessible flooring, constructed as a floating floor
2
absorption for sound impact decoupling
3
base plate of the building
4
magnet foundation, constructed as bonded screed
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...