40
ID: 182463115
Planning details
Measurements of the individual filter plates
Table 6.6: Measurements of the individual filter plates
6.2.1.1
Electronic filter plate
Position, alignment, and assembly
▪
In a horizontal position approx. 5 cm below the ceiling of the Faraday cage.
▪
The ideal position is on the wall adjacent to the technical room, as close as possible to the
service end of the magnet. If another position is selected, unnecessarily long cable lengths
must be anticipated.
▪
The electronic filter plate is assembled in an installation frame.
6.2.1.2
Magnet filter plate
Position, alignment, and assembly
▪
In a horizontal position approx. 5 cm below the ceiling of the Faraday cage. The magnet filter
plate is assembled directly beside the electronic filter plate.
▪
The ideal position is on the wall adjacent to the technical room, as close as possible to the
service end of the magnet. If another position is selected, unnecessarily long cable lengths
must be anticipated.
▪
The magnet filter plate is assembled in an installation frame.
6.2.1.3
Filter plate for anesthetic gas exhaust
Position, alignment, and assembly
▪
Beneath the ceiling of the Faraday cage
▪
The ideal position is directly above the service end of the magnet, above the CCM
If there are no detailed drawings for the filter plates described below, the measurements to be
used are those found in the following table.
Filter plate
H
B
D
B1
H1
B2
H2
B3
H3
C
[mm]
Electronics
410
1100
7
40
40
170
110
10
10
22
Magnet
220
450
7
45
50
120
120
7
7
22
Anesthetic gas ex-
traction
195
195
7
10
10
87,5
87,5
10
10
22
Magnet exhaust vent
260
260
6,5
10
10
80
80
10
10
22
MRI CryoProbe
TM
270
340
5
12,5
12,5
105 122,5 12,5
12,5
22
In-vivo
200
280
7
10
10
130
90
10
10
22
Ventilation
See chapter
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...