14.3 Checking battery performance
Please refer to the following steps to check the battery performance:
Disconnect the monitor from the patient and stop all monitoring or measurement.
Connect DC power to the monitor, and charge the battery for more than 4 hours
uninterruptedly.
Disconnect the DC power and power the monitor with battery until the monitor is turned
off.
Battery duration reflects the battery performance.
If the battery operating time is significantly shorter than the time stated in specifications,
please contact our service personnel for replacing the battery.
WARNING
•
Do not open the battery compartment. Only the qualified service personnel
authorized by the manufacturer can open the battery compartment and replace
or check the battery, and when it needs replacement, the replacement should be
same model with established battery.
14.4 Battery recycling
If the battery has visible damage or cannot store power, it should be replaced and recycled
properly. Follow the appropriate regulations to dispose of used batteries.
WARNING
•
Do not disassemble the battery, throw it in fire, or short-circuit it. Battery fire,
explosion and leakage may lead to personal injury. Do not touch the leaking
battery with bare hand directly.