BT-770 Operation manual
89
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
12
TEMP
12.1 Overview
The monitor has two temperature measurement channels; the temperature sensor will
measure the body temperature, and calculate the difference between the body temperature
data.
The monitor is designed for defibrillation proof, so the monitor operates normally after
defibrillation.
12.2 Safety information
WARNING
Before monitoring, check if the probe cable is normal. Unplug the temperature
probe cable from the jack, the screen will display “TEMP1 Sensor Off” and
“TEMP2 Sensor Off” prompt and make an alarm sound.
Calibrate the temperature measuring instrument at least once every two years
(or according to hospital procedures). When calibration is required, please
contact Bistos.
12.3 Measurement steps
Please refer to the following steps:
1.
Turn on the monitor and check if it works normally.
2.
Select the appropriate temperature probe according to the patient category and
measurement needs.
3.
Insert the probe lead wire into the temperature probe interface.
4.
Attach the probe to the patient properly.
5.
Make sure that the alarm settings apply to the patient.
When measuring body temperature, temperature probe can be attached to body surface such
as the neck, armpits, ears and other locations.