6.5 Alarm setup
6.5.1 Setting the alarm delay time
To limit alarm of continuous measurement parameter, you can set the alarm delay time. If the
alarm condition disappears during the delay period, the monitor will not generate an alarm. In
“Alarm Param” menu, select “Alarm Delay” time and “ST Alarm Delay” time.
The specific operation is as follows:
Select
[Settings] smart hotkey or press
[Setting] key on the monitor panel →
“Settings”.
Select “User Maintenance >>” → enter the password and confirm → “User Maintenance”
menu.
Select “Alarm Param >>” → “Alarm Param” menu.
Select “Alarm Delay”, and set the alarm delay time as needed:
“Off”: Turn off the alarm delay.
“1s” / “2s” / “3s” / “4s” / “5s” / “6s” / “7s” / “8s”: Alarm delay time is 1 sec, 2 sec, 3
sec, 4 sec, 5 sec, 6 sec, 7 sec or 8 sec. By default, the alarm delay time is 4 seconds.
Select “ST Alarm Delay”, and set the ST alarm delay time as needed.
“Off”: ST alarm delay is off
“10s” / “20s” / “30s” / “45s” / “1min” / “2min” / “3min”: ST alarm delay time is 10
sec, 20 sec, 30 sec, 45 sec, 1 min, 2 min or 3 min. By default, the ST alarm delay time
is 20 seconds.
6.5.2 Setting the alarm reminder signal and alarm reminder interval
The alarm reminder signal can be turned on or off. When the alarm is off and the alarm sound
is off, and then the alarm reminder signal is on. You can set the alarm reminder interval as
needed: “1min” / “2 min” / “3 min”.
The specific operation is as follows:
Select
【
Settings
】
Smart Hotkey→ “Settings” .
Select “User Maintenance >>” →enter the password and confirm → “User Maintenance”
menu.
Select “Alarm Param >>” → “Alarm Param” menu.
Select “Alarm Reminder Signal”, and set the alarm reminder signal as needed: