BT-770 Operation manual
54
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
6.4.3 Alarm pause
Press the
[Pause] smart hotkey or
[Alarm pause] key on the monitor panel to
temporarily stop the alarm of the monitor in the following steps:
[Pause]smart hotkey will appear magnified and reverse colored
icon.
The light alarm and audible alarm of the physiological alarms are suspended, and the
alarm information is not displayed.
The remaining time of alarm pause is displayed in the physiological alarm area.
Alarm parameters and upper / lower limit stop flashing.
The audible alarm and light alarm of technical alarms are suspended, but the alarm
message is still displayed.
After the alarm pause is finished, the monitor will automatically cancel the alarm pause state.
During the alarm pause, you can also press
[Pause] smart hotkey or
[Alarm pause] key
on the monitor panel to cancel the alarm pause manually.
You can set the alarm pause time as follows:
Select
[Settings]Smart Hotkey or press
[Setting] key on the monitor panel →
“
Settings
”
.
Select “User Maintenance >>” → enter the password and confirm →“User Maintenance”
menu.
Select “Alarm Param >>”→“ Alarm Param” menu.
Select “Alarm Pause Time”, and set the alarm pause time.
“1min” /“2min” /“3min” /“4min” /“5min” /“10min” /“15min” “Permanent”. By
default, the alarm pause time is 2 minutes.
“Permanent” means alarm off.
It is recommended that the SpO2 alarm pause time shall not more than two minutes.