6.4 Alarm states
In addition to the above alarm modes, you can also set the monitor to the following three
alarm states as needed, and display different alarm icons on the screen:
Alarm Reset
Alarm sound off
Alarm pause
Alarm off
6.4.1 Alarm reset
Select
button, and you can temporarily turn off the alarm sound of currently occurring
physiological alarms of the monitor, but the alarm information is still retained. For technical
alarms, clear the alarm state, display alarm prompt information, the alarm state icon area
displays the icon. When a new physiological alarm or technical alarm occurs, the alarm
reset is automatically canceled.
6.4.2 Alarm sound off
The alarm sound can be turned off through the following operations:
Select
[Settings] smart hotkey→ “Settings”.
Select “User Maintenance >>”→enter the password and confirm →“User Maintenance”
menu.
Select “Alarm Param >>”→“ Alarm Param” menu.
Set “Minimum Alarm Volume” to “0”. “Minimum Alarm Volume” range from 0 to 9, the
default value is 1.
Select
【
Volume
】
smart hotkey →“Volume Setup” menu.
Set “Alarm Volume” to “0”.
When the alarm sound is turned off, the alarm state area on the screen shows the
icon.
If “Minimum Alarm Volume” is larger than 0, the system will cancel alarm sound off state.
WARNING
When the alarm is off, and the alarm reminder signal is on, the system will have
alarm reminder tone.