12.4 Measurement requirements
The normal measuring range of monitor is 0
℃
~50
℃
, and the accuracy is consistent in this
range.
The environmental temperature range for body temperature measuring is 5
℃
~40
℃
. Get the
right temperatures for the shortest measurement time is 40s, and the measuring interval is 1s.
WARNING
Please measure the body temperature in the specified environment temperature
range, or else it may be dangerous.
12.5 Temperature display
The monitor can display the body temperature of two channels (T1 and T2) and the alarm
limits, difference between the two temperature (TD) and temperature units. Select Temp
parameter area and open the [Temp Setup] menu.
Temperature display area is as shown below:
Figure 12-1: TEMP parameter display
WARNING
The operator, prior to use, need to check the compatibility of the probe and
thermometer. If the temperature value displayed by the monitor has significant
difference from the body temperature under normal condition, please check if
the probe resistance of the monitor matches the resistance set in the monitor
system; if not, please replace a probe with appropriate resistance or adjust the
monitor and select the appropriate resistance. Incompatible probe will affect the
critical properties.