7.3.5 Checking the pacemaker
Before ECG monitoring, it is very important to set the pace maker state of the patient properly.
If the patient has a pacemaker, set “Pace Maker” to “Yes”, and the icon
displays in the
patient information area. When the system detects a pacing signal, the “
┃
”symbol will be
marked in the top of the ECG wave.
You can change the pacing state in the following method:
Select the patient information area to pop up the “Patient Info” menu, or select the
[Pat. Sat] smart hotkey and select “Patient Info” menu.
Select “Yes” / “No” for “Pace Maker” as needed, indicating that the patient with or
without pacemaker.
Diagnostic, Monitor, Surgery will not affect rejection of pacemaker pulses.
WARNING
•
For patients with pacemakers, the cardio tachometer may count the pacemaker
pulse in case of a cardiac arrest or arrhythmias. Never rely solely on the cardio
tachometer alarm. Closely monitor the patients with pacemaker. For the