BT-770 Operation manual
115
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
20
Default parameter configuration
This chapter lists the important factory default settings of different departments in monitor
configuration mode. Users can not change the default configuration, but can modify the
settings as required and save as user-defined configuration.
Module
Option
Module defaults
Adult
Pediatric
ECG
Alarm level
Mid
Mid
Alarm record
Off
Off
Lead type
5-lead
5-lead
Calculation channel
Auto
Auto
Power frequency suppression
On
On
Alarm limits
50~120 on
75~160 on
ST
segment
analysis
ST segment analysis
Off
Off
Alarm level
Mid
Mid
Alarm record
Off
Off
Alarm limits
-0.2~0.2 on
-0.2~0.2 on
Arrhythmia
analysis
Alarm level
Mid
Mid
Alarm record
Off
Off
Alarm limits
0~10 on
0~10 on
ARR
alarm
settings
Alarm switch
On
On
Alarm level
Mid
Mid
Alarm record
Off
Off
Gain
x1
x1
Wave velocity
25.0mm/s
25.0mm/s
Filter mode
Monitor
Monitor
Wave color
Green
Green
Wave style
Color scale
Color scale
NIBP
Alarm level
Mid
Mid
Alarm record
Off
Off
Pressure unit
mmHg
mmHg
Measurement mode
Adult
Pediatric
Interval
Manual
Manual
Display color
White
White
Pre-inflation value
150
100
Systolic blood pressure limit
90~160 on
70~120 on
Mean blood pressure limit
60~110 on
50~90 on
Diastolic blood pressure limit
50~90 on
40~70 on
SpO
2
Alarm level
Mid
Mid
Alarm record
Off
Off
Alarm limits
90~100 on
90~100 on
Wave velocity
25.0
25.0
Wave color
Cyan
Cyan
Wave style
Line
Line
RESP
Alarm level
Mid
Mid
Alarm record
Off
Off
Apnea alarm
20 sec
20 sec