6.7 Manual event
In the patient monitoring process, some events may have an impact on the patient, resulting in changes
of some monitoring waves or parameters. In order to assist in the analysis of these effects, you can
manually record these events through the
[Event] smart hotkey, and then view it in the event review,
refer to15.4 Event Review for detailed operation.
6.8 Alarm record
When the monitor’s machine alarm system is powered down, all alarm records are not saved.
Physiological alarm can store 200 alarm records, if full of 200, the latest alarm records will replace the
beginning of the record;
Technical alarm can store 100 alarm records, if full of 100, the latest alarm records will replace the
beginning of the record.