BT-770 Operation manual
87
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
11.5 NIBP display
NIBP measurement has no waveform display, and only displays NIBP measurement results in
the parameter area, as shown in Fig. 11-1. The figure below is for reference only. The graphics
displayed on the monitor may be slightly different.
Figure 11-1: NIBP parameter display
11.6 Setting inflation pressure
If necessary, you can manually set the initial cuff inflation pressure as follows.
Select the NIBP parameter area → “NIBP Setup” menu;
Select “Initial Pressure”, and set the appropriate cuff pressure value. When the patient is
adult, the pressure can be select from “140”, ”160” , ”180”. The default cuff pressure
value is “160”.
Select “Initial pressure”, and set the appropriate cuff pressure value. When the patient is
pediatric, the pressure can be select from “140”, ”160”. The default cuff pressure value is
“140”.
11.7 NIBP reset
Select NIBP parameter area → “NIBP Setup” menu → Select “Reset”, and restore the inflation
pressure of the blood pressure pump to currently configured initial settings. When the blood
pressure pump is not working properly, but no warning is given, you can reset the blood
pressure pump, and automatically restores the blood pressure pump.