6.4.4 Alarm off
As shown in 6.4.3, if the “Alarm Pause Time” is set to “Permanent”, press the
[Pause] smart
hotkey or
[Alarm pause] key on the monitor panel, and the monitor will turn off the alarm.
In this case, except the alarm prompt characteristics maintained in alarm pause state:
[Pause] smart hotkey will appear magnified
icon.
The physiological alarm area displays “Alarm Pause”.
You can press the
[Pause] smart hotkey again to manually cancel the alarm off.
If the monitor is in the alarm state of suspension or high priority technical alarm is triggered,
the alarm and the alarm off pause are automatically canceled.
WARNING
When the alarm volume is set to ‘0’ or the alarm pause time is set to permanent,
the monitor does not sound an alarm when an alarm occurs. Therefore, the
operator should use this feature carefully.