Patient information management
Connect the patient to the monitor, and the monitor will display and store the physiological
data of the patient, so the patient can be monitored without admitting the patient. However,
admitting the patient correctly is very important.
If the monitor has admitted the patient, it is recommended to operate the monitor to
discharge the current patient before connecting to (not admitted) the next patient. Otherwise,
the data of the previous patient will be stored in the data of the current patient.
WARNING
Whether the patient is admitted or not, the system will give a default value to
“Patient Type” and “Pace Maker”, “Patient Type” default “Adult”, “Pace Maker”
default “No”, and the user must confirm that the default value is appropriate for
the patient being monitored.
For patients with pacemakers, “Pace Maker” must be set to “Yes”. Otherwise, the
pacing pulse will be treated as normal QRS wave group, and the system is unable
to detect the alarm status of “ECG Signal weak”.
For patients without a pacemaker, “Pace Maker” must be set to “No”. Otherwise,
the system is unable to detect the arrhythmias (including PVCs count) related to
ventricular premature beats, and fails to perform ST segment analysis.
4.1 Patient setup menu
You can manage the patient through the “Patient” menu. To enter “Patient” menu, operate as
follows:
Select
[Settings]Smart Hotkey or press
[Setting] key on the monitor panel →
“
Settings
”
→ “Patient >>” → “Patient” menu;
Or
Select
[Pat. Set] Smart Hotkey to enter “Patient” menu, as shown in Fig. 4-1.