BT-770 Operation manual
76
P/N : 770-ENG-OPM-EUR-D03
Bistos Co., Ltd.
2018.05
Do not put the sensor on limbs with arterial duct or intravenous tube.
Do not intertwine electrosurgical equipment cable with the sensor cable.
Avoid using the monitor and sensors while using the NMR equipment, in order to
avoid severe burns to the patient as a result of induced currents.
During long time continuous monitoring of a patient, check the position of SpO2
sensor once every 2 hours, and move properly when the skin changes or every
four hours. Some patients may require more frequent inspection, such as
patients with perfusion disorders or sensitive skin, because persistent and
prolonged monitoring may increase unpredictable skin changes, such as
allergies, redness, blistering or pressure necrosis.
When the measured pulse rate is not complete, then the "---".
Before using, verify compatibility between the monitor, probe and cable,
otherwise it may cause injury to the patient.
Functional testers cannot be used to evaluate the accuracy of pulse oximetry and
pulse oximetry.
SpO2 low alarm limit cannot be less than 85.
NOTE
Do not put the oxygen probe and blood pressure cuff on the same limb, because
blood flow occlusion during blood pressure measurement will affect the SpO
2
readings.
The monitor cannot be used to verify the accuracy of SpO
2
probe and SpO
2
equipment.