BIOQUELL PROTEQ USER MANUAL -EN
TD092-O&M-001
REVISION 5 DRAFT
Page 41 of 58
Alarm
No.
Alarm Name
Alarm
Description
Action
User Action
5
Vaporiser 1
temperature high
Vaporizer
temperature too high
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
6
Vaporiser 2
temperature low
Vaporizer not
reaching temperature
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
7
Vaporiser 2
temperature high
Vaporizer
temperature too high
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
8
Delivery
temperature low
Heaters are not able
to achieve the
required air
temperature
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
10
Sterilant injection 1
fault
Vaporizer not
receiving Hydrogen
Peroxide
Unit will go into emergency
aeration
Aerate until room
reaches 1ppm and
contact service provider
11
Sterilant injection 2
fault
Vaporizer not
receiving Hydrogen
Peroxide
Unit will go into emergency
aeration
Aerate until room
reaches 1ppm and
contact service provider
12
Pump rotation fault
Hydrogen peroxide
dosing pump is not
rotating
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm. Remove
bottle and contact
service provider.
13
Distribution fan 1.1
fault
Distribution fan 1 has
stopped working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
14
Distribution fan 1.2
fault
Distribution fan 1 has
stopped working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
15
Distribution fan 2.1
fault
Distribution fan 2 has
stopped working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
16
Distribution fan 2.2
fault
Distribution fan 2 has
stopped working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
17
Distribution fan 3.1
fault
Distribution fan 3 has
stopped working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
18
Distribution fan 3.2
fault
Distribution fan 3 has
stopped working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
19
Distribution fan 4.1
fault
Distribution fan 4 has
stopped working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider
20
Distribution fan 4.2
fault
Distribution fan 4 has
stopped working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm and
contact service provider