BIOQUELL PROTEQ USER MANUAL -EN
TD092-O&M-001
REVISION 5 DRAFT
Page 28 of 58
Description
Units Limits
Explanation
Injection in Dwell
Check or
Uncheck
Check to allow the injection of H
2
O
2
during dwell phase. Timed cycle
only.
Aeration Parametric
Check or
Uncheck
Check allows the aeration phase to
finish after the aeration time has
been completed and will continue, if
required, until concentration is below
1ppm. If unchecked the aeration will
run for the set time only.
CYCLE PARAMETERS
Gassing Injection
Rate*
g/min 4 to 16
4 to 12 (with
Flexi Adaptor)
4 to 10 (100V)
H
2
O
2
injection rate during gassing
phase.
H2O2 Mass to inject
g
10 to 4520
Mass of liquid H
2
O
2
solution
vaporised in gassing phase. (timed
cycle only, gassing parametric
disabled)
Room Volume
m
3
5 to 400
Room volume to be decontaminated
(gassing parametric enabled)
Dwell Injection Rate* g/min 4 to 16
4 to 12 (with
Flexi Adaptor)
4 to 10 (100V)
H
2
O
2
injection rate during dwell
phase.
Dwell Time
Sec
60 to 21600
Length of time in dwell phase
H2O2 Alert Level
ppm
0 to 1000
At the end of gassing dwell the
system checks that the peak PPM
reached thus far is greater than the
alert level. If not a warning alarm
message will be displayed and
printed.
Aeration VFC Delay
Sec
0 to 64800
Delay time before the volt free cycle
phase indicator switches from
gassing to aeration. Delay time must
be shorter than aeration time
Aeration Time
Sec
60 to 59940
Length of aeration phase if Aeration
Parametric is unchecked, and
minimum aeration time if checked.
+ Aeration Units
0 to 6
Number of additional wireless
aeration units which are required.
* These values must be set with care to ensure hydrogen peroxide does not
condense in the vaporiser. Seek advice from Bioquell.