Page 17 of 58
If a Flexi Adaptor is fitted the left hand light will illuminate green and the right
hand will be unlit. No other colours will be seen.
7.2
Installing Hydrogen Peroxide Bottles
The following instructions explaining how to install bottles must be followed; in
addition, there is an illustration of the procedure on the vaporizer module above
the bottle enclosure. Before loading a 950ml bottle ensure the insert is present,
see section 7.3.
WARNING:
Use Personal Protective Equipment when handling hydrogen
peroxide bottles. As a minimum, wear suitable gloves and goggles.
Press the bottom right hand side of the flap
under the bottle storage compartment.
The flap will spring-up.
Remove a pair of gloves and goggles. Put them
on before handling the bottle of hydrogen
peroxide.
Loading new bottles:
STEP 1:
To access the bottles push the silver button at
the centre of the black plunger. The plunger will
be released and will move up.
STEP 2:
Obtain a new bottle and check the outer bag for
any signs of liquid. If liquid is present, do not
use the bottle. Treat the liquid as hydrogen
peroxide and take the necessary precautions. If
no liquid is present, remove the bottle from the
bag, and remove the lid carefully by turning
counter clockwise. Retain the lid.