BIOQUELL PROTEQ USER MANUAL -EN
TD092-O&M-001
REVISION 5 DRAFT
Page 43 of 58
Alarm
No.
Alarm Name
Alarm
Description
Action
User Action
32
Aeration unit 3
main fan failure
(if present)
Main aeration fan has
stopped working
Will prevent a cycle from
starting. If in cycle prior to
gassing will stop the cycle,
otherwise will go to
emergency aeration
Aerate until room
reaches 1ppm, and
contact service provider
33
Aeration unit 3
distribution fan 1
failure
(if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
34
Aeration unit 3
distribution fan 2
failure
(if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
35
Aeration unit 4
main fan failure
(if present)
Main aeration fan has
stopped working
Will prevent a cycle from
starting. If in cycle prior to
gassing will stop the cycle,
otherwise will go to
emergency aeration
Aerate until room
reaches 1ppm, and
contact service provider
36
Aeration unit 4
distribution fan 1
failure
(if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
37
Aeration unit 4
distribution fan 2
failure
(if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
38
Aeration unit 5
main fan failure
(if present)
Main aeration fan has
stopped working
Will prevent a cycle from
starting. If in cycle prior to
gassing will stop the cycle,
otherwise will go to
emergency aeration
Aerate until room
reaches 1ppm, and
contact service provider
39
Aeration unit 5
distribution fan 1
failure
(if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
40
Aeration unit 5
distribution fan 2
failure
(if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
41
Aeration unit 6
main fan failure
(if present)
Main aeration fan has
stopped working
Will prevent a cycle from
starting. If in cycle prior to
gassing will stop the cycle,
otherwise will go to
emergency aeration
Aerate until room
reaches 1ppm, and
contact service provider
42
Aeration unit 6
distribution fan 1
failure
(if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider