BIOQUELL PROTEQ USER MANUAL -EN
TD092-O&M-001
REVISION 5 DRAFT
Page 45 of 58
Alarm
No.
Alarm Name
Alarm
Description
Action
User Action
57
Hydrogen Peroxide
Sensor Fault
The H2O2 sensor is
reading a low value
to do a fault
Will prevent a cycle from
starting. If in cycle prior to
gassing will stop the cycle,
otherwise will go to
emergency aeration
Faulty EMM – contact
service provider
58
Aeration unit 1
distribution fan 3
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
59
Aeration unit 1
distribution fan 4
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
60
Aeration unit 2
distribution fan 3
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
61
Aeration unit 2
distribution fan 4
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
62
Invalid Cycle Recipe
Cycle requires the
Sensor pack option,
OR cycle has
injection rate higher
than allowed
Will prevent a cycle from
starting.
Delete cycle or edit
cycle parameters. OR
reduce injection rate to
12g/min for Flex
Adaptor or 10g/min for
100V
63
Decon Fan1 Fault
Fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Contact service
provider
64
Decon Fan2 Fault
Fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Contact service
provider
65
Aeration unit 3
distribution fan 3
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
66
Aeration unit 3
distribution fan 4
(flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
67
Aeration unit 4
distribution fan 3
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider