BIOQUELL PROTEQ USER MANUAL -EN
TD092-O&M-001
REVISION 5 DRAFT
Page 20 of 58
8
OPERATION
8.1
Running a Cycle
Check that the low level hydrogen peroxide sensor is
working. To start remove the cap over the sensor if
present and then hold down the green button. The
display will count down from 3 and then start. Place
the sensor in the recess on the lectern. If the sensor
is not working, do not run a cycle. The sensor may
be stored on the lectern, see section 4.1.2
Ensure the door is sealed, this is likely to require
taping. A roll of tape is supplied on the lectern.
WARNING –
Ensure all possible openings
and conduits into the enclosure are
properly sealed. If large leaks occur and
high concentration Hydrogen Peroxide
escapes into areas where people are
present it can cause serious harm. Use
tape provided to seal door.
Press icon in top right of screen to
login, if using local logins use the
down arrow to select user name
from the list and enter password –
touching the data entry box a
keyboard will appear.
If using network login (Windows)
then ensure that an ethernet cable
is connected between the lectern
and the local network, and enter
user name and password using the
keyboard which appears when the
data entry box is touched. Press
.
When logged IN the User Name will
be displayed in the top right of the
screen. The Login period will expire
after 5 minutes of inactivity if not in
cycle, after which it will be
necessary to log on again.