Alarm
No.
Alarm Name
Alarm
Description
Action
User Action
68
Aeration unit 4
distribution fan 4
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
69
Aeration unit 5
distribution fan 3
(flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
70
Aeration unit 5
distribution fan 4
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
71
Aeration unit 6
distribution fan 3
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
72
Aeration unit 6
distribution fan 4
failure (flat cat only
option, if present)
Aeration distribution
fan has stopped
working
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, then re-
run cycle with other
aeration units and
contact service provider
73
Insufficient
Aeration Units for
Loaded Cycle
Cycle requires more
aeration units than
are on the wireless
network
Will prevent a cycle from
starting.
Ensure the correct
quantity of aeration
units are available,
they are all powered up
and paired on the
wireless network. Try
re-pairing the network
74
Base unit
Communications
lost in cycle*
One of the ProteQ
vaporiser base units
is no longer
communicating with
the Multi-Generator
Controller
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, check
power to the base
units, unpair all base
units and re-pair.
75
Number of base
units incorrect for
loaded cycle*
The required number
of ProteQ vaporiser
base units for the
Multi-Generator
Controller cycle are
not on the network
Will prevent a cycle from
starting.
Ensure the correct
number of base units
are on the network, if
necessary unpair all
base units and re-pair.
76
Number of
vaporiser units in
cycle incorrect*
The required number
of ProteQ vaporiser
base units for the
Multi-Generator
Controller cycle are
no longer on the
network
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Aerate until room
reaches 1ppm, check
power to the base
units, unpair all base
units and re-pair.
77
Remote Stop*
A ProteQ vaporiser
base unit has been
stopped by the Multi-
Generator Controller
Will stop cycle prior to
gassing stage, otherwise will
go into emergency aeration.
Solve issue with the
ProteQ base unit which
has the alarm which
initiated the other units
to be stopped