52
www.bioneer.com
BQ-042-101-12
Revision : 0 (2018-07-05
Exicycler
™
384
Calibration using ExiCfg
The calibration must be performed prior to operating
Exicycler
™ 384 Real-Time Quantitative Thermal
Block. You can either transfer proper calibration data associated with a serial number from the
installation CD to the computer or perform the calibration by an authorized engineer of Bioneer. Install
the Operation and Analysis Software for the calibration data transfer to the computer. For a reliable
operation of the
Exicycler
™ 384, calibrating
Exicycler
™ 384 with a Fluorescence test kit is
recommended. Calibration must be done when moving
Exicycler
™ 384 to a different location or when
changing a light source lamp. A periodical calibration of every 6 months helps to maintain the optimum
condition of
Exicycler
™ 384.
Use the ExiCfg program to calibrate
Exicycler
™ 384. The ExiCfg calibrates as well as diagnoses
systemical condition of
Exicycler
™ 384. Make sure that you are fully trained for handling the ExiCfg
before starting calibration. Otherwise, authorized staff is only allowed to manage the ExiCfg. Mis-use of
the ExiCfg by unauthorized staff may cause serious damage to
Exicycler
™ 384.
The calibration is carried out in 4 steps, Warming up the lamp > Mask calibration > Background
calibration > Multi-Channel calibration.
Mask Calibration
1)
Turn on
Exicycler
™ 384 by pressing the POWER
button.
NOTE:
Make sure the status LED is blinking in green after the self-diagnosis prior to starting the Fluorescence
Check.
2)
Go to ‘C:\Exicycler384\Run’ and double click
ExiCfg.exe.
POWER button
Summary of Contents for Exicycler 384
Page 4: ...Exicycler 384...
Page 7: ...2 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Before Starting...
Page 11: ...6 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Safety Instructions...
Page 15: ...10 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Introduction...
Page 52: ...47 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Performance Maintenance...
Page 70: ...65 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Troubleshooting...