Turning On and Self-diagnosis
1)
Turn on
Exicycler
™ 384 by switching on the main power switch in the rear of the instrument.
When power is supplied properly, a status LED turns blue.
2)
Press the POWER button for a second to start self-diagnosis.
3)
When the status LED is blinking green, you can either open or close a main door by pressing a
DOOR button for a second. The status LED is blinking in purple when the door is open and is back
to green when the door is closed.
Power Off Power On Door Open Standby
IMPORTANT:
The status LED starts blinking green after two short beeps when the self-diagnosis is complete.
Exicycler
™ 384 is now ready to operate. The status LED will turn red when the self-diagnosis fails or
excessive physical shock is applied from outside. (e.g. Make sure that there are no objects placed in
front of the door) Make sure that the status LED is blinking green before operating
Exicycler
™ 384 to
prevent mis-operation or damage. A result of the self-diagnosis is provided under ‘ExiConfig’.
※ Status LED color
a.
Red: Error
b.
Blue: Standby
c.
Green: Ready for running
d.
Yellow: Self test
e.
Cyan: Running
f.
Purple: Door open
g.
White: Pause
NOTE:
You can stop the operation here and turn off the
Exicycler
™ 384 by pressing the POWER button for a
second when it is blinking in green like Step 3.
Do not place any obstacle in front of the system and do not load a plate or take it out when the door
is moving. It can cause serious damage and lead to severe injury.
Summary of Contents for Exicycler 384
Page 4: ...Exicycler 384...
Page 7: ...2 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Before Starting...
Page 11: ...6 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Safety Instructions...
Page 15: ...10 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Introduction...
Page 52: ...47 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Performance Maintenance...
Page 70: ...65 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Troubleshooting...