56
www.bioneer.com
BQ-042-101-12
Revision : 0 (2018-07-05
Exicycler
™
384
Background Calibration
1)
Press the DOOR button to open the door.
NOTE:
This step is omissible when performing the Background Calibration followed by the Mask Calibration.
2)
Place the Background Fluorescence test plate on
the block.
IMPORTANT:
It is important to load the Fluorescence test plate in the right position. A misplaced plate may cause
damage or technical problems.
3)
Click Background Calibration button.
4)
Click Start in the ‘Background Calibration’ window.
NOTE:
The Background Calibration takes approximately 10 minutes depending on the condition of
Exicycler
™
384.
POWER button