69
www.bioneer.com
BQ-042-101-12
Revision : 0 (2018-07-05
Exicycler
™
384
Error message when running
Exicycler™
384
software
(continued).
9.
Fail to load calibration information will appear if the computer
fails to read the calibration data from
Exicycler™
384. Go to C:
\Exicycler384 to check if the calibration folder and file exist.
10.
Melting Protocol can't be in Cycle Region will appear if the
melting step is inserted within a thermal cycling.
11.
Need to replace the Lamp will appear if the lamp intensity drops
below 60% of the normal lamp intensity. Replace the lamp.
12.
No matching calibration data with Machine. Can't Run! will
appear if you click Run button when
Exicycler™
384 ID and the
calibration data ID do not match.
13.
No matching calibration data with machine will appear if the
Exicycler™
384 ID and the calibration data ID do not match.
14.
Please insert any protocol before inserting Goto Protocol will
appear if the Goto step is inserted in the first line of a protocol
file.
15.
Please insert Data name will appear if you have not entered a
name for data prior to PCR running.
16.
Probe Information was changed, do you want to cancel this? will
appear if you click the Cancel button without saving the edited
probe Information.
17.
Probe using same filter is assigned will appear if you assign more
than 2 probe options within the same wavelength range.
18.
Ramping rate option Must be inserted between Incubation steps
will appear if you insert a ramping rate option in a wrong
position of the protocol file.
19.
Set Probe Name will appear if you have not entered a new name
for the custom probe in the Add Probe window.
20.
UserName was already in User List will appear when entering an
existing name for User set-up.
21.
You must select Dye and Quencher will display if the dye and
quencher set up has not been selected in the Add Probe
window.
Summary of Contents for Exicycler 384
Page 4: ...Exicycler 384...
Page 7: ...2 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Before Starting...
Page 11: ...6 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Safety Instructions...
Page 15: ...10 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Introduction...
Page 52: ...47 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Performance Maintenance...
Page 70: ...65 www bioneer com BQ 042 101 12 Revision 0 2018 07 05 Exicycler 384 Troubleshooting...