28
www.bioneer.com
BQ-042-101-12
Revision : 0 (2018-07-05
Exicycler
™
384
Troubleshooting
1.
Appropriate calibration data does not exist
1)
CD provided with
Exicycler
™ 384 has Calibration data for
each device. Calibration data have to be copied into
C:\Exicycler384\Run folder in order for a normal
function. If Calibration folder does not exist on this path,
following message will be displayed.
2)
Click OK to change the path.
3)
You should assign the path of the Calibration folder to be
copied automatically. Click Open and you will see the
following browser.
4)
After assigning the Calibration folder (ex. E: \Calibration),
click OK.
5)
Copy Calibration window displays the path of Calibration
folder and serial number of your device. If the serial
number matches with the connected
Exicycler
™ 384
system, click Copy button on the right hand side of
window.
6)
Please wait for a few minutes to complete.