NOTE:
An approximate calibration time for each dye is 5 minutes. Put the plate back into its packaging and
return it to the freezer.
High temperature
Wait at least 5 seconds before taking out the Fluorescence test plate from the 384-well block. The
Fluorescence test plate is heated to a temperature of 100°C and may cause a burn.
9)
Prepare the Fluorescence Test plate with a dye X (i.e.
“SYBR Green”) when ‘Attach the plate named ‘X’ (i.e.
“SYBR Green”) on the bath’ message appears.
10)
Place the Multi-Channel Fluorescence Test plate on the
block.
11)
Click Continue and repeat step 9 through 10.
12)
When the Multi-channel Calibration is complete, click
Close in the ‘Multi-Channel Calibration’ window.