Maintenance
Conditional maintenance
6
7
8
9
10
TEMPO
®
Reading Station
User’s Manual
9-7
C o n d i t i o n a l m a i n t e n a n c e
DANGER! Before any maintenance operations, the instruments and equipment
must be disconnected from the power source.
After disconnection, wait for 1 minute before performing maintenance
operations.
When removing a card which is jammed in the TEMPO
®
Reader, it is
obligatory to wear gloves to protect against biological hazards.
To open the access cover (captive screw), use the screwdriver provided
with the TEMPO
®
Reader. It is imperative to close the access cover once
the maintenance operation is over.
Summary of co ndi tional mai ntenance operations
♦
Unlocking the
TEMPO
®
Reader
door manually ........................................ 9-8
♦
Replacing fuses ......................................................................................... 9-9
♦
Configuring the Reading Station barcode reader..................................... 9-11
♦
Removing a jammed card from the
TEMPO
®
Reader
............................ 9-14
Summary of Contents for TEMPO
Page 2: ......
Page 6: ......
Page 26: ......
Page 30: ......
Page 38: ......
Page 52: ......
Page 56: ......
Page 58: ......
Page 110: ......
Page 212: ......
Page 228: ......
Page 250: ......
Page 252: ......
Page 254: ......
Page 255: ...TEMPO Reading Station User s Manual NOTES 1 Notes ...
Page 256: ...Index NOTES 2 TEMPO Reading Station User s Manual ...
Page 257: ...Index TEMPO Reading Station User s Manual NOTES 3 ...
Page 258: ...The logo is a registered and protected trademark of bioMérieux sa or one of its subsidiaries ...