Maintenance
Systematic preventive maintenance
9-4
TEMPO
®
Reading Station
User’s Manual
Monthl y maintenance
Cl eaning an d decontami n ation
Note:
cleaning and decontamination operations can also be performed:
−
in case of accidental contamination,
−
if you consider decontamination is necessary.
Duration of check
< 1 hour
Checked by
a laboratory technologist
Spare part(s) and consumables
Not applicable
Tools and laboratory
equipment
Lint free paper
Maintenance products
A chlorinated solution diluted to 3°
chlorometry
70° alcohol
T E MP O
®
Reader
•
Clean the external surfaces of the
TEMPO
®
Reader
using a disinfectant
solution (70° alcohol or a chlorinated solution diluted to 3° chlorometry ).
•
Wipe the external surfaces of the
TEMPO
®
Reader
using moistened lint
free paper (to remove any residue of chlorinated solution).
•
Dry the external surfaces of the
TEMPO
®
Reader
using lint free paper.
Internal surfaces
The internal surfaces of the
TEMPO
®
Reader
must not be cleaned by the user.
If the internal surfaces require cleaning, call bioMérieux
®
or your local
bioMérieux
®
representative.
Summary of Contents for TEMPO
Page 2: ......
Page 6: ......
Page 26: ......
Page 30: ......
Page 38: ......
Page 52: ......
Page 56: ......
Page 58: ......
Page 110: ......
Page 212: ......
Page 228: ......
Page 250: ......
Page 252: ......
Page 254: ......
Page 255: ...TEMPO Reading Station User s Manual NOTES 1 Notes ...
Page 256: ...Index NOTES 2 TEMPO Reading Station User s Manual ...
Page 257: ...Index TEMPO Reading Station User s Manual NOTES 3 ...
Page 258: ...The logo is a registered and protected trademark of bioMérieux sa or one of its subsidiaries ...