Figure 4.57. Sealed Source Screen with Isotope to be Edited
3.
Touch to highlight each field to edit on the Sealed Source Setup screen.
4.
Make the appropriate edits, touching <Accept> to save each one. When all the required
edits have been made, touch <Save >. The system saves the changes made and displays a
note stating that sealed source changes have been saved.
5.
Touch <Accept> to return to the Sealed Source screen.
6.
Touch <Back> to return to the Sealed Source List screen.
To Delete A Sealed Source:
1.
At the Sealed Sources List screen, touch to highlight the sealed source to delete.
2.
Touch <Delete>. The system will note that the sealed source selected will be deleted and
prompt "Are you sure?"
Figure 4.58. Delete Sealed Source Prompt
3.
Touch <Accept> to delete the sealed source selected and return to the Sealed Source List, or
<Cancel> to abort the deletion and return to the Sealed Source List without making
changes.
To Mark A Sealed Source As Default:
1.
At the Sealed Sources List screen, touch to highlight any listed sealed source and touch
<Mark as Default> to mark it as the default. The default symbol ( ) is displayed to the left of
the selected sealed source denoting that it has been marked as the default. This will be the
first source displayed when entering the constancy and accuracy screens.