Figure 7.38. Future Dose Calculation Screen Illustrating Result of User Entering
the Dose Concentration.
9.
Touch <Accept> to return to the Future Dose Calculation screen.
10.
Place the drawn dose into the detector to confirm the activity drawn in the syringe.
11.
Touch <Print Labels> to generate a label if a label printer is connected.
NOTE:
If drawing additional doses, enter the administration time and dose ordered for
another dose and the system calculates the requirements for that dose.
12.
Touch <Clear> to remove the <Dose Concentration> and <Dose Ordered> data from the
screen.
13.
Touch <Measure>to advance to the Dose Calibrator Measure screen.
14.
Touch <Zero Background>. If there has been no activity in the chamber, a “Completed”
message is immediately displayed and the system returns to the Future Dose Calculation
screen.
15.
Touch <Back> to return to the Nuclear Pharmacy screen.
16.
Touch <Home> to return to the Main Atomlab Menu.