3.
Touch <Back> to return to the Linearity History screen.
To Delete Data:
1.
Touch <Delete Data>. The Delete Stored Test Data confirmation screen is
displayed:
Figure 6.83. Delete Stored Test Data Screen.
2.
Touch <Accept> to delete the current test.
3.
Touch <Cancel> to return to the Decay Test History screen.
To View Details:
1.
Touch a specific line on the Decay Test History screen to highlight the information.
2.
Touch <View Details>. The Decay Test View Details screen is displayed.
Figure 6.84. Decay Test View Details Screen.
3.
Use the <
▲
> and <
▼
> to toggle between dates.
4.
Touch <Print> in the upper right of the screen to print the displayed test details.
5.
Touch <Delete> to delete the test details.
6.
Touch <View Decay Plots> to advance to the Linearity Decay Plots screen and display a
plot of the selected linearity data.
7.
Touch <Back> to return to the Decay Test History screen.